Skip to main content
All six acquisition cards

View 1Subcostal 4-Chamber

Probe 2 cm below the xiphoid, marker to the patient's left — both ventricles, both atria, tricuspid and mitral valves, no LVOT.

Free to print and laminate for your own machines — no sign-up. Steps follow the EASy MAP methods paper.

Set your printer to Letter, landscape and background graphics on. One card per sheet — trim and laminate for the machine.

1

Subcostal 4-Chamber

View 1 · probe position A · the cardiac window

EASy MAP

Machine setup

  1. Sector (phased-array / cardiac) probe — the same probe for all six views.
  2. Adult Cardiac preset, B-mode.
  3. Marker indicator on the upper right of the screen (cardiology convention).
  4. Depth 21 cm.
  5. Frequency at the probe's lowest setting (typically 2 MHz).
  6. TGC on; increase far gain to fill in the far field.

Acquire

  1. Gel the transducer. Patient supine if tolerated and practical.
  2. Place the probe 2 cm inferior to the xiphoid process, marker toward the patient's left (3 o'clock), overhand grip.
  3. Trap a small skin fold under the probe; significant inward pressure is often needed. Attend to comfort in awake or lightly sedated patients.
  4. Angle the probe tail caudally and toward the patient's right, coming slightly out of the axial plane — this aims the beam cephalad and toward the patient's left.
  5. A segment of liver should appear in the superficial image: the left lobe is the acoustic window that carries the beam to the heart.

Pitfalls

  • Not enough gel, or not enough probe pressure — the commonest cause of a poor novice image.
  • Gripping the probe too far from the emitter, or too tightly, kills fine control.
  • Subcostal alone can underestimate RV size and miss the cardiac midsegment and apex.
  • In very low-flow states or on mechanical support, swirling heterogeneous echogenicity in the ventricles is stasis, not a gain problem.
  • A poor-quality cardiac view (≤ 6) is not usable for cardiac decisions — the IVC, lung and pleural findings may still be actionable.

Optimize

  • At 21 cm the sector should already extend slightly beyond the posterior LV wall, so the whole pericardium is included — if the heart still doesn't fit, add depth.
  • Set gain so blood filling the chambers is black; some visible turbulent flow is acceptable.
  • Off-axis chambers → rotate counterclockwise. Chambers cut off or off-centre → rock right or left.
  • LVOT in view → you are in a 5-chamber plane; lift the tail to drop it out.

Confirm before you record

A correctly acquired view shows both ventricles, both atria, and the tricuspid and mitral valves — but not the left ventricular outflow tract. Record a 4 s clip (longer if the ventricular rate is < 30 bpm).

Read it in this order

  • Pericardium — non-trivial effusion (> 10 mm, through systole and diastole). Tamponade is a clinical diagnosis, supported by atrial collapse in systole or ventricular collapse in diastole.
  • RV size and function — an RV larger than the LV (RV/LV > 1) is severe dilation; look for a diastolic free wall > 1 cm or an enlarged RA as signs of chronicity.
  • Septum — thinning, rupture, hypertrophy; shift toward the LV if the RV is dilated.
  • LV size and function — severe dilation, marked wall thickening, or minimal systolic cavity reduction with poor mitral annular motion.
  • Then integrate with the IVC, the lung profile and the clinical picture — not before.

12-point quality score

Score each 0–2Total
Pericardium · RV size · RV function10–12 good
Septum · LV size · LV function7–9 adequate
0 = not assessable, 1 = usable, 2 = clear≤ 6 poor
EASy MAP image acquisition · Howell-Clark J, Bronshteyn YS, Pustavoitau A, Convissar DL, Bughrara N. J Vis Exp. 2025;(218):e67531. Free for educational use — not a substitute for clinical judgment.